电子试验机的标定方法

发布时间 2022-11-04     发布人: admin     浏览次数:408

Select "Force Calibration" in the pull-down menu under the "Calibration" menu in the menu of the electronic universal testing machine, and a dialog box will appear. At this time, enter the password to enter the calibration interface. In the first operation, the password is 123456, and you can directly select "OK". Then you will see the calibration interface on the right side of the interface. Press the [Calibration Settings...] button, and the test parameter modification interface will appear. Then select the force calibration settings, First enter the range, then select the number of calibration points, and then enter the calibration value you want. Press the "Start Calibration" button to enter the next interface, and you can start the calibration operation 

Calibration method: Place the mouse on the point you want to calibrate in advance, and then you will see that the selected point to be calibrated has a black dotted box, which represents the selected point. When the point to be calibrated is approaching, press the left button of the mouse directly when calibration is needed. Then, when calibrating the next point, and so on. After all points are calibrated, the system will prompt that calibration is complete 摩擦磨损试验机 端淬试验机 拉力试验机 材料试验机 电子试验机 扭转试验机

在电子试验机的菜单中选中"标定"菜单下的下拉菜单中的"力值标定",就会出现一个对话框,此时输入密码就可以进入标定界面,在第一次运行时密码:123456,直接选择"确定"就可以了,然后你就会在界面的右边看到的标定界面,先按下[标定设置…]按钮,出现试验参数修改的界面,然后选择力值标定设置,先输入量程,然后选择标定点数,然后再输入你要的标定的值就可以了.按下"开始标定"按钮后,就进入下一个界面了,就可以进行标定操作了.   

标定方法:将鼠标提前放在你要标定的点上,此时会看到被选中的要标定的点有一个黑色的虚线框,那个就代表选中的点.在要标定的点快要到的时候,需要标定时直接按下鼠标的左键.接着标定下一个点时,依此类推.所有的点都标定完后系统会提示,标定完成.